Function
|
1. The air exchange rate shall not be less than 180 times per hour.
2. The load-bearing capacity of the laminated board design shall not be less than 70kg/m2.
3. The operating status and fault alarm device of the fan should have a response time of no more than 10 seconds.
According to the requirements of GB/T 9068, a single reagent cabinet should operate normally with a noise level not exceeding 68dB (A).
5. Real time temperature and humidity environment monitoring system, displaying temperature and humidity, and setting alarm parameters.
6. Fan monitoring: Fan system failure alarm, online adjustable fan speed to ensure the required air volume for different chemical operations.
7. Filter saturation alarm system: The product is equipped with double-layer filters and dual VOC probes. One probe monitors indoor air quality and the other monitors filter saturation status. The filter sets a saturation alarm value, which will sound an alarm if it exceeds the range. When the concentration exceeds the set value for a long time, the filter needs to be replaced.
8. PSC fan imported from the United States, with 24V voltage, stable performance, ultra quiet, and no spark static electricity.
9. Efficient filtration system, arranged and distributed according to particle size, following ASTM standards, effectively targeting acidic gases and organic gases, with strong adsorption capacity. High efficiency HEPA filters are used for particles, with a filtration efficiency of 99.995% for particles larger than 0.3um.
10. Supports SIM card 4G communication
11. It has the function of managing the chemicals operated, and can support the display of a list of no less than 100 chemicals, including serial numbers, chemical names, quantities, units, etc; On the device side, operations such as adding, deleting, and changing the list can be performed; Synchronize data to the cloud backend server; The WeChat client can view a list of currently used chemicals.
12. Remote management can be performed through the WeChat client, where historical VOC data and temperature and humidity environmental data can be viewed and presented in a curve format; Historical alarm records can be consulted, and fans, lighting, and alarm lights can be remotely controlled.
13. It can provide API interfaces and support third-party platform docking functions.
|